Specification |
Lunatic instrument specifications |
| Dimensions | 37 cm W x 46 cm D x 33 cm H; 21 kg |
| Operating voltage | 24 VDC, 30 W (max) |
| Connection | USB, TCP/IP (Service, SiLA) |
| Light source | Xenon flash lamp |
| Detectors | UV/Vis polychromatic spectrophotometer |
| Approval | CE, FCC, CSA |
| Wavelength range | 230-750 nm |
| Wavelength accuracy | ≤400 nm: ±1 nm; ≥400 nm: ±2 nm |
| Spectral resolution | Better than 2 nm (toluene in hexane) |
| Absorbance precision (1 cm quartz cuvette) | <1 OD: ±0.005 OD st dev | 1-2 OD: ±0.5% CV |
| Absorbance accuracy (1 cm quartz cuvette) | <1 OD: ±0.01 OD | 1-2 OD: ±1% |
Lunatic plate specifications | High Lunatic Plate | Lunatic Plate |
| Application area | High concentration samples, typically proteins | Low concentration samples, typically nucleic acids |
| Sample retention time | Up to 2 hours | Up to 2 hours |
| Recommended sample volume | 2 µL | 2 µL |
| Pathlength(s) | 0.1 mm & 0.7 mm path | 0.5 mm path |
| Measurement time for full plate | ~10 minutes | ~5 minutes |
| Measurement range: |
| OD 10 mm | 0.03–275 OD 10 mm | 0.03–40 OD 10 mm |
| ng/µL dsDNA | 1.5–13750 ng/µL | 1.5–2000 ng/µL |
| mg/mL ave protein mix | 0.03–275 mg/mL | 0.03–40 mg/mL |
| Absorbance precision (10 mm pathlength) | <1 OD: ±0.01 OD st dev | <1 OD: ±0.01 OD st dev |
| 1-200 OD: ±1% CV | >1 OD: ±1.5% CV |
| Absorbance accuracy (10 mm pathlength) | <1 OD: ±0.02 OD | <1 OD: ±0.02 OD |
| 1-200 OD: ±2% | >1 OD: ±4% |
| Dimensions | 8.55 cm W x 12.8 cm D x 1.21 cm H |
| Samples per plate | 96 (12 x 8 microplate format) |